ChipFind - документация

Электронный компонент: NLP110-9917

Скачать:  PDF   ZIP
LOW TO MEDIUM POWER AC/DC POWER SUPPLIES 80-110W AC/DC Universal Input Switch Mode Power Supplies
1
2 YEAR WARRANTY
NLP110 Medical Series
S i n g l e a n d t r i p l e o u t p u t
Provides low voltage outputs
Compliance to EN61000-3-2 (Power Factor = 0.98)
Smallest industry standard package 6.5 x 3.0 x 1.26 inches
90VAC to 264VAC universal input range
UL, cUL, and VDE safety approvals and CE-Marked to MDD
Overvoltage and short circuit protection
VDE0871-A, EN55022-A, EN55011-A conducted noise
VDE0871-A, EN55022-A, EN55011-A radiated noise
Meets all applicable and relevant immunity standards
EN61000-4-2,4,5 and 6
The NLP110 Medical is a series of 110W AC/DC open-frame power supplies in a
3" x 6.5" x 1.26" package. The NLP110 Medical is power factor corrected to meet EN61000-3-2
and is available in single and multiple output models. The multiple output models have a capability
that is unique to the power supply market - they can deliver equal amounts of current on multiple
low voltage outputs to support designs that will span the industry transition from 5V logic to 3.3V
logic. The NLP110 Medical is practically the same size as a 65 Watt 3" x 5" product, e.g. NLP65
Medical - the one difference is that the footprint is a mere 1.5 inches longer. The NLP110 Medical
series, with full medical safety approval per UL2601-1, C22.2 No. 601.1-M90, VDE0750,
IEC60601-1 and EN60601-1 and CE marked, greatly accelerating design-in time and reducing
system compliance costs.
All specifications are typical at nominal input, full load at 25C unless otherwise stated
EMC CHARACTERISTICS (continued)
Surge
EN61000-4-5
Level 3
Fast transients
EN61000-4-4
Level 3
Radiated immunity
EN61000-4-3
Level 3
Conducted immunity
EN61000-4-6
Level 3
GENERAL SPECIFICATIONS
Hold-up time
120VAC, 60Hz
16.7ms @ 110W
Efficiency
120VAC, 80W (-9605)
75% min.
Isolation voltage
Input/output
4000VAC
Input/chassis
1500VAC
Approvals and
EN60601-1, IEC60601-1,VDE0750,
standards
UL2601-1, C22.2 No. 601.1-M90
Weight
283g (10oz)
MTBF (@ 25C)
MIL-HDBK-217F
220,000 hours min.
ENVIRONMENTAL SPECIFICATIONS
(9)
Thermal performance
Operating ambient,
0C to +50C
(See derating curve)
Non-operating
-40C to +85C
50C to 70C, ambient,
Derate to
convection cooled
50% load
0C to 50C, ambient,
80W
convection cooled
0C to 50C ambient,
110W
150LFM forced air
Peak (0C to +50C, 60s)
(See Note 3)
Relative humidity
Non-condensing
5% to 95% RH
Altitude
Operating
10,000 feet max.
Non-operating
30,000 feet max.
Vibration (See Note 8)
5Hz to 500Hz
2.4G rms peak
Shock
per MIL-STD-810E
516.4 Part IV
OUTPUT SPECIFICATIONS
Total regulation
Main output
2.0%
(Line and load)
Auxiliary outputs
5.0%
Rise time
At turn-on
1.0s, max.
Transient response
Main output
5.0% or 250mV
75% to 100%
max. dev., 1ms max.
step at 0.1A/s
recovery to 1%
Temperature coefficient
0.02%/C
Overvoltage protection
Main outputs
125%, 10%
Short circuit protection
Cyclic operation
Continuous
Minimum output current
Single and multiple
See table
INPUT SPECIFICATIONS
Input voltage range
Universal input
90 to 264VAC
(See Note 6)
Input frequency range
47Hz to 63Hz
Input surge current
120VAC
18A max.
(cold start)
230VAC
35A max.
Safety ground
230VAC, 50Hz
<100A
leakage current
Input current
120VAC @ 80W
0.9A rms
120VAC @ 110W
1.3A rms
230VAC @ 80W
0.48A rms
230VAC @ 110W
0.7A rms
Input fuse
UL/IEC127
F3.15A H, 250VAC
EMC CHARACTERISTICS
(12)
Conducted emissions
EN55022, FCC part 15
Level A
Radiated emissions
EN55022, FCC part 15
Level A
Harmonic current
EN61000-3-2
Compliant
emission correction
ESD air
EN61000-4-2
Level 3
ESD contact
EN61000-4-2
Level 3
SPECIFICATIONS
File Name: NLP110M.PDF Rev: 13 Dec 2002
For the most current data and application support visit www.artesyn.com/powergroup/products.htm
Notes
1
Free air convection.
Multiple output units: maximum continuous output power not to exceed
80W.
For -9993; I
3.3V
= 13A max.; I
5.0V
= 13A max.; I
3.3V
+ I
5.0V
16A.
For -9994; I
3.3V
= 13A max.; I
2.5V
= 13A max.; I
3.3V
+ I
2.5V
16A
For -9995; I
3.3V
= 13A max.; I
12V
= 6.5A max.; I
3.3V
+ I
12V
16A.
For -9908; I
5.0V
= 13A max.; I
12V
= 6.5A max.; I
5.0V
+ I
12V
16A.
Single output units: maximum continuous output power not to exceed;
75W on 9905; 76.8W on 9912,-9924, and 9917.
2
150LFM forced air cooling from L4 side.
Multiple output units: maximum continuous output power not to exceed
105W.
For -9993; I
3.3V
= 16A max.; I
5.0V
= 16A max.; I
3.3V
+ I
5.0V
20A.
For -9994; I
3.3V
= 16A max.; I
2.5V
= 16A max.; I
3.3V
+ I
2.5V
20A
For -9995; I
3.3V
= 16A max.; I
12V
= 8.5A max.; I
3.3V
+ I
12V
20A.
For -9908; I
5.0V
= 16A max.; I
12V
= 8.5A max.; I
5.0V
+ I
12V
20A.
Single output units: maximum continuous output power not to exceed
110W for all models.
3
300LFM forced air cooling from L4 side.
Multiple output units: maximum continuous output power not to exceed
110W.
For -9993; I
3.3V
= 20A max.; I
5.0V
= 18A max.; I
3.3V
+ I
5.0V
22A.
For -9994; I
3.3V
= 20A max.; I
2.5V
= 20A max.; I
3.3V
+ I
2.5V
22A
For -9995; I
3.3V
= 20A max.; I
12V
= 8.5A max.; I
3.3V
+ I
12V
22A.
For -9908; I
5.0V
= 20A max.; I
12V
= 8.5A max.; I
5.0V
+ I
12V
22A.
Single output units: maximum continuous output power not to exceed
110W for all models.
4
Peak output current lasting less than 30 seconds with duty cycle less
than 5%. During peak loading, output voltage may exceed total
regulation limits.
5
Figure is peak-to-peak for convection power rating. Output noise
measurements are made across a 20MHz bandwidth using a 6" twisted
pair, terminated with a 10F electrolytic capacitor and a 0.1F ceramic
capacitor.
N
6
Minimum load required for correct start-up and operation on single
outputs and on main output of multiple versions. Failure to observe
minimum load on main output will not allow the supply to start-up
correctly. Some electronic test loads have a large delay time before they
start drawing current even though the voltage from the supply is present.
During this time delay, there is no load on the output and as a result, the
supply cannot start-up properly and maintain its correct output voltage.
In these instances, a dummy resistive load across the output may be
necessary to load the output of the supply until the test load can function
correctly and draw the intended minimum load. Minimum load required
on auxiliary outputs to maintain regulation.
7
For models NLP110-9908 and NLP110-9995, the 12V output is floating.
For -12V output, pin 11 on J2 has to be connected to Return making pin
12 the -12V output
8
Three orthogonal axes, random vibration 10 minutes for each axes, 2.4G
rms 5Hz to 500Hz.
9
For optimum reliability, no part of the heatsink should exceed 110
o
C, and
no semiconductor case temperature should exceed 120
o
C.
10
CAUTION: Allow a minimum of 1 second after disconnecting line power
when making thermal measurements.
11
This product is only for inclusion by professional installers within other
equipment and must not be operated as a stand alone product.
12
The EMI specifications reference measurements made with the power
supply mounted on a grounded metal sheet extending 1 inch beyond
each edge, using an unshielded cable. No external filtering required
during conducted emissions testing but some applications may require
additional filtering to achieve system compliance. A line choke, (AC input
cords looped twice through an EMI suppression toroid) was used during
radiated emissions testing. Considerable radiated testing in 1U six-sided
boxes has shown that units can meet level B in typical systems.
Application support is available from the factory to assist with EMI
compliance.
13
All models require a minimum mounting stand-off of 6.35mm (0.25
inches) in the end use product.
NLP110 Medical Series
S i n g l e a n d t r i p l e o u t p u t
LOW TO MEDIUM POWER AC/DC POWER SUPPLIES 80-110W AC/DC Universal Input Switch Mode Power Supplies
2
www.artesyn.com
File Name: NLP110M.PDF Rev: 13 Dec 2002
International Safety Standard Approvals
VDE0750/IEC60601-1 and EN60601-1
UL2601-1 and C22.2 No. 601.1-M90.
OUTPUT
OUTPUT CURRENT
RIPPLE
(5)
TOTAL
MODEL
VOLTAGE
MIN
(6)
MAX
(1)
150LFM
(2)
300LFM
(3)
PEAK
(4)
REGULATION
NUMBERS
+5V
0.5A
15A
22A
22A
22A
50mV
2.0%
NLP110-9905
+12V
0.3A
6.4A
9.2A
9.2A
11.5A
120mV
2.0%
NLP110-9912
+48V
0A
1.6A
2.3A
2.3A
2.5A
240mV
2.0%
NLP110-9917
+24V
0.2A
3.2A
4.6A
4.6A
6.0A
240mV
2.0%
NLP110-9924
+5V (A)
0.5A
13A
16A
18A
18A
50mV
2.0%
NLP110-9993
+3.3V (B)
0.2A
13A
16A
20A
20A
50mV
2.0%
+12V
0A
0.65A
1.0A
1A
1.0A
120mV
5.0%
+3.3V (A)
0.5A
13A
16A
20A
22A
50mV
2.0%
NLP110-9994
+2.5V (B)
0.1A
13A
16A
20A
22A
50mV
2.0%
+12V
0A
0.65A
1.0A
1A
1.0A
120mV
5.0%
+12V (A)
0.2A
6.5A
8.5A
8.5A
10A
1200mV
2.0%
NLP110-9995
(7)
+3.3V (B)
0.5A
13A
16A
20A
22A
50mV
2.0%
-12V
0A
0.65A
1.0A
1A
1.0A
120mV
5.0%
+12V (A)
0.2A
6.5A
8.5A
8.5A
10A
120mV
2.0%
NLP110-9908
(7)
+5V (B)
0.2A
13A
16A
18A
22A
50mV
2.0%
-12V
0A
0.65A
1.0A
1A
1.0A
120mV
5.0%
DERATING CURVE
Output Power (Watts)
0
C 10C20C30C40C50C60C 70C
40W
80W
110W
150 LFM FORCED
AIR COOLING
55W
NATURAL
CONVECTION
COOLING
Input and output connectors
Mating connectors
AC (J1) connector type
AC (J1) mating connector type
Molex 26-60-4030 or equivalent.
Molex 09-50-3031 or equivalent with Molex
08-50-0105 or equivalent crimp terminals.
DC (J2) connector type
DC (J2) mating connector type
12 position Molex Spox type 26-48-1125
Molex Spox type 26-03-3121 and
or equivalent.
contact 08-52-0113.
Sense (J3) connector type
Sense (J3) mating connector type
Molex 22-23-2021 or equivalent.
Molex 22-01-3027 and contact
08-50-0113.
Earth (J4) connector type
Earth (J4) mating connector type
Male 0.250 quick disconnect.
Molex 90028.
Sense (J5) connector type
Sense (J5) mating connector type
Leoco 2421P02H000.
Leoco 2420S020000 and contact
2453TPB00V1.
0.23
(5.842)
6.50
(165.10)
1
T1 (REF.)
J4
J1
6.050
(153.670)
2.550
(64.770)
ALL DIMENSIONS IN INCHES (mm)
J2
J3
1
1
2
12
3.00
(76.20)
F1
J5
2
1
0.23
(5.842)
!
3
MAXIMUM COMPONENT
HEIGHT IS 1.26 (32.00)
OUTPUT PIN CONNECTIONS
J2
SINGLE
TRIPLE
Pin 1
No Connection
V (B)
Pin 2
No Connection
V (B)
Pin 3
No Connection
V (B)
Pin 4
Return
Return
Pin 5
Return
Return
Pin 6
Return
Return
Pin 7
Return
Return
Pin 8
V (A)
V (A)
Pin 9
V (A)
V (A)
Pin 10
V (A)
V (A)
Pin 11
No Connection
V (C)
Pin 12
No Connection
V (C) Return
(7)
Recommended direction
for forced air relative to
power supply orientation
shown below.
1
Optimum.
2
Very good.
3
Good.
4
Not recommended.
2
3
4
1
INPUT
PIN CONNECTIONS
J1
Pin 1
AC Neutral
Pin 2
No Connection
Pin 3
AC Line
J3
Pin 1
V (A) Sense +
Pin 2
V (A) Sense -
J4
Pin 1
Safety Earth
J5
Pin 1
V (B) Sense +
Pin 2
V (B) Sense -
Data Sheet Artesyn Technologies 2002
The information and specifications contained in this data sheet are believed to be correct at time of publication. However, Artesyn Technologies accepts no responsibility for consequences arising
from printing errors or inaccuracies. Specifications are subject to change without notice. No rights under any patent accompany the sale of any such product(s) or information contained herein.
For the most current data and application support visit www.artesyn.com/powergroup/products.htm
NLP110 Medical Series
S i n g l e a n d t r i p l e o u t p u t
LOW TO MEDIUM POWER AC/DC POWER SUPPLIES 80-110W AC/DC Universal Input Switch Mode Power Supplies
3
www.artesyn.com
File Name: NLP110M.PDF Rev: 13 Dec 2002
Mechanical Notes
A
All dimensions are in inches (mm).